Samuel Lesko, senior manager for optical and tribology applications, Bruker Nano Surfaces and Metrology
The concept of measuring surface roughness originated nearly a century ago as a means to prevent uncertainty and disputes between manufacturers and buyers. Now it has become a common identifier used throughout industry for validating manufacturing processes, confirming adherence to both internal and regulatory specifications, and guaranteeing quality and performance of end products. Subjective judgements of quality based on naked eye observation or fingertouch feel of surfaces has steadily been replaced by unbiased metrics and well-defined formulas.
The first parameter developed for these endeavours was mean or average roughness (commonly referred to as Ra), which, for a number of reasons, is still a primary reference parameter used today. First, mean roughness is easy to work out, even in an analogic way, which not only was important in early implementation across a variety of industries, but also makes it a convenient and quick method for current characterisation. Second, the Ra parameter is a robust calculation that averages outlier data and provides constant results irrespective of the roughness pattern (see figure 1). This is critically important to not only assist with a wide variety of industrial manufacturing processes but also provide a solid baseline for process improvement.
Figure 1: Illustration of mean roughness calculation.
Adoption of Ra as a key parameter to qualify surface roughness relies on defined standard samples (standards) of the desired profile measurement. These are relatively easy to manufacture for a single line profile and are commonly used to assess if a roughness measurement system is properly calibrated. This ensures that the Ra value given for a specific surface is linked back to a well-established reference value. Such standards also help to achieve a common reference across multi-industrial sites and provide tool-to-tool correlation in multi-metrology systems. While these standards were originally designed for the single-line measurements of stylus-based profilers, they are also able to confirm that both contact and non-contact areal-based profilers acquire the correct results.
This article discusses the use of mean roughness measurements with white light interferometry (WLI) optical profilers. Spatial filters are explained as well as some of the normative standard requirements from ASME B46.1:20191, ISO 13565–1:19962, ISO 4287:19973 and JIS B 0671–1:20024. Main technical reasons for WLI selection are covered as well as the advantages and areas of applicability of the full areal measurement standard from ISO/DIS 25178–25.
Roughness and spatial filtering
In manufacturing, products are defined through key dimensions summarised in technical drawings. Roughness is one specific critical parameter that defines how much a surface measurement deviates from a specified shape or form, with height variation within the millimetre lateral range. Larger fluctuation in topography is part of another parameter, namely waviness. The surface of a product is the aggregate of form, shape, waviness and roughness, all defined to a desired volume. In this respect, roughness can only be worked out after the proper selection of spatial components, excluding form, shape and waviness (see figure 2). Roughness measurements always include step, where shape is removed, either directly via a physical skid (as with some stylus profilers) or via post-processing (e.g., high-order polynomial fitting and/or spatial filtering).
Figure 2: Determining roughness for a cylindrical part.
In the latter case, long-range topography fluctuations or equivalent low spatial frequencies are leveraged out through a high-pass filter, which only allows fast topological variation to pass through. Vice-versa, short rapid variation on a profile more often indicates the presence of noise, which should not be considered for a roughness measurement. In such cases, the high spatial frequencies are excluded through a low-pass filter (see figure 3). Thus, raw measurement profiles go through band-pass filters for lower and higher spatial limits. In ASME B46.1 and ISO 4287 norms, those boundaries are defined as λc and λs cut-off parameters.
Figure 3: Spatial filter definition with associated derived parameters and resulting profiles.
While filtering is an essential part of measuring roughness, it should be kept in mind that filters can also distort measured topography. For example, the traditional use of resistor-capacitor (RC) filters was well-known for triggering edge effects that required the use of exclusion bands at the beginning and end of profiles to remove spurious data. RC filters are also prone to distort topography around sudden height variations, such as peaks or pits. Therefore, this filter has been mostly replaced in current usage by more reliable filters, such as Gaussian and spline-phase-compensated filters. The more recent use of areal roughness has initiated the need for even more advanced filters, such as the robust Gaussian filters6, which are part of the ISO 25178 norm. These filters provide a unique advantage in avoiding edge effects (see figure 4). This allows a roughness calculation by effective removal of waviness across the entire measurement field while accurately capturing fine variations in topography.
Figure 4: Plot showing the advantage of a spatial filter.
Profile versus areal measurements
For almost a century, stylus-based profiles have successfully captured mean roughness and assessed the quality of parts. Common turning, milling and other CNC machining processes all leave well-defined direction where roughness occurs. In such cases, dragging the stylus perpendicularly to the main machining traces provides reliable and precise measurement of a surface’s texture. This type of measurement is now referred to as a 1D type, where height (Z) is expressed versus scanning length (X).
Over the last two decades, however, requirements for increased manufacturing efficiency and both energy and cost savings have driven a higher level of complexity for surface metrology. Surfaces are now engineered to serve specific purposes, making the roughness parameter an even more critical parameter for most precision-engineered parts. Surfaces are textured in multiple directions to increase performance (e.g., with a lower coefficient of friction) or to improve lifetime or wettability (to name just a few). All these processes are challenging to assess with a single-line profile, making multi-site measurements and a better statistical approach mandatory (see figure 5).
Three types of surfaces showing different applicability for profiler or areal topography measurements: (A) profile roughness standard; (B) cross-hatch texture on bore cylinder; and (C) defects on an optical window.
With these rising limitations in pure 1D roughness measurements, full areal roughness metrology, known as 2D measurements, has become the most common means of characterising surface roughness. These measurements plot vertical height versus X and Y directions. The first techniques available were also based on stylus profiler capabilities, combining multiple adjacent lines into an area. The main drawbacks for this approach were the long measurement times, usually an hour or more to achieve high lateral resolution in both directions, and the inherent fluctuation between each line due to mechanical drift. In the 90s, several groups succeeded in utilising optical microscopes to measure topography, based on interferometry techniques7.This enabled fast (seconds to minutes) non-contact roughness measurements. Since then, areal roughness measurement techniques have expanded to include confocal, focus variation and digital microscope.
These techniques all capture the full field of view and work out a dedicated height position for each pixel of an image. These techniques are now referred to as 2D (area). It should be noted that there are also 3D techniques that measure full volume, including re-entrant and porous surface (e.g., confocal X-ray tomography). However, these techniques excel at different parameters than surface roughness. With a 2D or areal measurement, it becomes much easier to capture surface texture in all directions, as well as to spot any random defect along a surface. It also captures a larger field, which makes measurements more representative of overall surface texture as well as more robust through the higher amount of statistical data available.
In modern industry, 1D and 2D measurement techniques co-exist. For standard manufacturing processes that leave a single texture orientation, mean roughness can easily be captured through a few profiles. If a texture has two or more orientations, both stylus profiling and areal measurements are valid; with stylus profiling requiring multiple lines, more time and further statistical analysis to accurately present the surface. Finally, in the case of a random and/or complex engineered texture, optical areal measurement becomes mandatory to properly evaluate the surface.
White light interferometry profiler
Due to the increased complexity of much manufacturing today, non-contact areal profilers are widely utilised to measure roughness in both R&D and production environments. Many design variations exist, but they commonly share a dedicated rigid structure that serves as a platform for high-end objectives and digital cameras (see figure 6 as an example). They also share the fact that lateral and vertical resolutions are objective dependent, which has led to the common use of highly resolutive objectives with short working distances for best vertical resolution. A major exception to this trend is WLI-based profilers, where vertical resolution not only becomes independent from the objective but also uniquely reaches sub-nanometre levels.
Figure 6: White light interferometry (WLI) optical implementation with display of different Moiré pattern versus illumination.
A WLI profiler utilises interferometric objectives that reveal the sample surface via a moiré pattern only when proper focus is reached. As the depth of field for the moiré presence does not exceed ±0.5 µm due to the limited coherence length of white light illumination, the focal plane can be easily worked out through finding the maxima within a couple of nanometres. This sharp determination of focal plane relies entirely on moiré and is independent from the objective, which guarantees nanometre precision, even with low-magnification objectives (e.g., 1, 2.5 or 5x). As shown in figure 7, there are several positive consequences of this approach:
- long-working-distance objectives can be used without compromising vertical resolution to access specific or recessed locations on a complex part;
- ease of use is increased with the extra safety margin between objective and surface, as well as with the ability to target challenging locations;
- a mirror can be inserted along the focusing beam, deflecting the optical path to measure vertical walls with higher precision;
- if low lateral resolution is needed, a single acquisition at low magnification covers a wide range (100 mm2), making rapid detection of defects possible or permitting high-throughput flatness control;
- stitching can be used to combine high lateral resolution over even wider areas; and
- metrology assessment and budget allocation become easier since all objectives have the same vertical precision.
Figure 7: Attributes of WLI metrology.
A WLI-based profiler not only complies and is listed as an appropriate technique for areal norm ISO 25178-204:2013 but, due to its unique vertical resolution, is used for artifact calibration work by major reference metrology laboratories such as the National Institute of Standards and Technology (NIST), National Physical Laboratory (NPL) and Physikalisch-Technische Bundesanstalt (PTB). Based on this work, some advanced WLI profiler designs utilise a direct reference to a stabilised HeNe laser for automatic and self-calibration purposes8.
Roughness calculations
The ISO 11562-1997, ASME B46.1 and JIS B 0632:2001 standards clearly define measurement conditions, as well as filtering, for surface profiles. Prior to the roughness calculation, the raw profile is corrected for shape. Then it goes through a low-pass filter determined by cut-off λs to obtain the primary profile, P. Roughness parameters are worked out after further high-pass filters, defined by cut-off λc, to remove waviness. All R parameters are derived from this roughness filtered profile. The filters are defined by mathematical function; most often Gaussian or spline-phase corrected. For periodic profiles (e.g., milling, turning, etc.), both cut-off parameters are set to expected mean roughness for random surfaces or with spacing between peaks (see table).
From the cut-off values, standards derive the exact measurement travel (Lt), which includes evaluation length (Lm) extended by pre- and post-lengths. In the case of a Gaussian filter, pre- and post-lengths correspond to λc/2 length to reduce influence of potential mechanical backlash at the start and end of a scan, as well as filter edge effects. This precaution is not necessary for modern optical profilers and robust Gaussian filters. The evaluation length corresponds to a series of five sample lengths; itself being equal to the λc cut-off value (see figure 8). Such stringent conditions are necessary not only for extraction of relevant information from the complete profile but also to ensure seamless comparison between different instruments. Mean roughness results directly correlate with selected cut-off values as well as with the spatial filter used. Any change of measurement parameters conversely modifies the roughness output: whenever inter-comparison between different measurement systems is engaged, all filtering parameters must be identical.
Figure 8. Details on total measurement length.
For areal measurement, ISO 25178-2 is the sole standard for roughness calculation. This norm leaves selection of cut-off values and evaluation length to the operator but dictates step-by-step filtering of the raw surface. Raw surface data acquired by the profiler first gets low-pass filtered (digital or spatial) to remove noise and outliers. The result is the primary surface value. Most common spatial filters are based on Gaussian process regression, or robust Gaussian process regression, which provides a better response on sharp transitions and has almost no effect on borders. The primary surface is further processed with shape removal to create the S-F* surface value. To remove waviness, an extra high-pass filter can be used to produce an S-L** surface value. Users can further crop borders to obtain the scale-limited surface and work out roughness parameters. In such conditions, the areal roughness standard does not distinguish parameters: they are all labelled as S parameters. ISO 25178, however, does require listing the whole post-processing chain prior to displaying parameters, such as cross-check control (see figure 9). The same applies to roughness specifications on technical drawings, where exact measurement conditions must be clearly labelled.
Figure 9: Step-by-step processing for areal roughness.
Measurement of roughness standard with WLI
One of the core applications of WLI-based profilers is measuring sub-micron roughness and surface texture. The technique’s unique sub-nanometre vertical resolution, combined with its large field of view and long-working distances, make it suitable for metrology of all precision-engineered parts, ranging from forged flat metal to complex/curved surfaces, such as gears or bore cylinders. In addition to this flexibility, the WLI optical profiler metrology capability benchmarks to certified roughness standards (see figure 10). Here, Bruker investigated nine standards from three different manufacturers, namely Halle, NPL and Rubert.
Figure 10: Topography rendered in 3D view of different roughness standards: (A) areal; (B-C) random profiles; and (D) sinusoidal profile.
The goal was to cover extensive lateral and vertical ranges to ascertain metrology capability across a wider number of applications (see figure 11). This approach also offers comprehensive assessment for linearity performance over the vertical range. One areal artifact was also part of the evaluation to represent a real case for this native areal measurement method.
Figure 11: Plot of nominal mean roughness versus required lateral resolution.
Technical considerations
Regular 1D roughness standards are primarily designed for stylus-based profilers. In that respect, lateral resolution capability does not depend on scan length but rather on aspect ratio and depth of the surface features. On the other hand, an optical profiler has inherent changes in lateral resolution whenever a low-magnification objective is in use. It is consequently important to select an objective that achieves a lateral resolution better than the λs cut-off. Operators must also ensure that camera sampling is at least twice the optical resolution to effectively achieve the expected lateral resolution. Otherwise, lateral resolution becomes limited by pixel size. Once the objective and zoom lens are properly selected, the evaluation length is ensured either by single acquisition or by the stitching of multiple fields of view.
Here, 5, 20, 50 and 115x objectives were used. The lowest objective (5x) was utilised for rougher surfaces that required less lateral resolution and longer evaluation length, while the highest objective (115x) was critical to resolve the finest patterns.
All data first underwent a 4th order polynomial removal before going through a Gaussian regression band-pass filter. Cut-offs were selected as per the ISO 11562-1997 norm and settings from the standard certificate. Final roughness extraction consisted of separating the areal image as a series of single-line profiles for which mean roughness (Ra) was worked out. Results not only showed the average mean roughness along all profiles, but also indicated fluctuation of the Ra value, one sigma deviation.
To address precision, measurements were repeated 30 times over each roughness standard in a fully static way. This follows the recommendation from the Guides to the expression of uncertainty in measurement (GUM)9. Coverage factor k=2 was chosen, representing 95 percent of results for a purely random Gaussian distribution.
Results
All mean roughness results are summarised in figure 12, plotting measured Ra value versus certified value in log-log scale. In assessing the quality and consistency of the WLI optical profiler, there is great correlation over two decades, from sub-100 nanometres to over a micron mean roughness. Error bars are represented for each result using ±2s dispersion from static repeatability. However, they are hardly visible since, on average, they are below 1 percent of the result. This emphasises how repeatable the WLI profiler is.
Figure 12: Measured mean roughness by WLI versus certified nominal value in log-log display.
Further validation of results relies on the comparison between measurement of the dispersion interval and the confidence interval from the standard certificate. Figure 13 illustrates different configurations and indicates whether a decision can be made from the data results. Since the dispersion range always lies within the uncertainty of the standard, all the results obtained with the WLI profiler are positively assessed. Data prove the accuracy of the WLI profiler measurements across the entire range of the available roughness standards.
Figure 13: Mean roughness results assessment versus certification.
Interestingly, the WLI profiler provides reliable data at both extremes of the tested spectrum. On the rougher side, a low-magnification (5x) objective delivers precise measurement. Robust algorithms for topography extraction from optical data, together with high-power illumination, provides a good ability to measure surfaces with rough, steady slopes. This extends the capability of the WLI profiler to tens of microns roughness, enabling the measurement of additive manufactured parts for instance. At the other extreme, sub-micron fine pitch can be clearly resolved by a WLI profiler, providing adequate lateral resolution. This can be seen by a direct comparison of WLI profiler measurements with those of Bruker’s Dimension Icon atomic force microscope (AFM). Figure 14 exhibits the correlation between section profiles made with the two techniques. A high numerical aperture 115x objective, together with sub-nanometre vertical resolution and an advanced super-resolution algorithm, expand the transfer function of the optical profiler beyond micron lateral size.
Figure 14: Profile comparison on sinusoidal standard between WLI optical profiler and atomic force microscope (AFM).
Conclusions
This article has provided a comprehensive overview on how surface roughness has evolved as a key manufacturing parameter, from technique capabilities to normative guidance on both profile and areal measurements. The WLI profiler measurement examples against certified roughness standards show perfect correlation with certified values. With a dispersion range below 1 percent and a precise mean value, the WLI profiler is capable of measuring mean roughness over two decades (from 3 µm to 40 nm). WLI profilers also have a proven ability to measure steady, rough slopes, as well as to achieve sub-micron lateral resolution, all while maintaining extremely precise vertical measurement. These factors indicate that WLI profiling will continue to play an integral role in the ever increasingly stringent R&D and manufacturing requirements for next generation industrial products.
Acknowledgements
Special thanks to Reinhard Danzl and Franz Helmi from Bruker Alicona (Graz, Austria) for samples and review.
Bruker Nano Surfaces and Metrology
*S-F surface is a surface filter obtained after applying an F-operator to the primary surface. An F-operator removes form from the primary surface.
**S-L surface is a surface optained after applying an L-filter to the S-F surface. An L-filter eliminates the largest scale elements from the surface (also known as a high-pass filter).
References1American Society of Mechanical Engineers (ASME) (2020). ASME B46.1:2019 Surface texture (surface roughness, waviness and lay).Available at: https://bit.ly/32UOvAR2International Standards Organisation (ISO) (1996). ISO 13565–1:1996 Geometrical product specifications (GPS)—surface texture: profile method; surfaces having stratified functional properties—part 1: filtering and general measurement conditions.Available at: https://bit.ly/3nvaU1i
3International Standards Organisation (ISO) (1997). ISO 4287:1997 Geometrical product specifications (GPS—surface texture: profile method—terms, definitions and surface texture parameters.Available at: https://bit.ly/3wj0XqJ
4JIS Standards (2002). JIS B 0671–1:2002 Geometrical product specifications (GPS)—surface texture: profile method, surfaces having stratified functional properties—part 1: filtering and general measurement conditions.Available at: https://bit.ly/338JF3f
5International Standards Organisation (ISO) (under development). ISO/DIS 25178–2 Geometrical product specifications (GPS)—surface texture: areal—part 2: terms, definitions and surface texture parameters.Available at: https://bit.ly/3gKgjjF
6Seewig, J. (2005). Linear and robust Gaussian regression filters. Journal of Physics, conference series 13 059, pp.254–257.
7Wyant, J. C. (1995). Computerised interferometric measurement of surface microstructure. Society of Photo-Optical Instrumentation Engineers (SPIE) proc. volume 2576, International Conference on Optical Fabrication and Testing, Tokyo, Japan, June 5–7.
Available at: https://bit.ly/3e0K6CK
8Novak, E. (2003). White-light optical profiler with integrated primary standard. IMEKO proc., XVII IMEKO World Congress, Metrology in the 3rd millennium, Dubrovnik, Croatia, June 22–28.
Available at: https://bit.ly/2QlQxr8
9Guides to the expression of uncertainty in measurement [web page]. Joint Committee for Guides in Metrology.
Available at: https://bit.ly/3hq6rvv